Recording a 12-lead ECG OSCE mark scheme, Year 3

Record a 12-lead ECG on a woman admitted after a dizzy spell, talked through to the examiner: who she is, where each of the ten electrodes goes, what the machine is set to, what you do when the first trace cannot be read, and whose name is on the paper.

Marked at

Year 3 is where this sheet starts: 27 lines and 32 marks, most of them on systematic approach and patient safety.

  • 1:00 + 8:00 reading, then the room
  • 5 examiner questions
  • Sat from Year 3
  • Free to practise, no account needed
  • Reviewed
AI OSCE station · UKReal screen
A live AI OSCE skills station from the UK library, peripheral cannulation: the candidate says each step and an examiner tells them what happens. The desk shows the kit on the trolley and the candidate's notes.
AI OSCE station · UKReal screen
The candidate's desk on a phone during an AI OSCE knee examination from the UK library: the instructions, and the observations the examiner has just handed over.

How do you want to practise?

Live with the AI1 credit, £3.67–£6Sit it live

Candidate instructions

The notice on the door

Mrs Brenda Ashworth, 72, was brought to the acute medical unit an hour ago after a dizzy spell while waiting for a bus. The doctor looking after her has asked for a 12-lead ECG. Please record it, talking the examiner through each thing you check, say and do, and why, including exactly where each electrode goes. There is nothing to touch: the examiner will tell you what you see and what the patient says. You have eight minutes. Towards the end the examiner will ask you some questions about recording an ECG.

1:00Read8:00In the roomBell
Ready when you are1:00

1 min reading, then 8 min in the station

On the trolley

  • ECG machine with a ten-wire patient cable
  • Disposable tab electrodes in a foil pack
  • Skin-cleansing wipes and gauze swabs
  • Single-use razor
  • Sheet to cover the patient
  • Alcohol hand gel
  • Pen
  • Sharps bin and clinical waste bag
Your notes from reading time

Saved in this browser only.

Year 3 is where this sheet starts: 27 lines and 32 marks, most of them on systematic approach and patient safety.

Year 3 standard. Records a readable trace from correctly placed electrodes on a correctly identified patient, deals with the poor first recording at the bedside, and checks the name on the paper without being asked.

2 critical lines. Miss one and the station is failed, however well the rest goes.

PreparationMarks
Identity and consent
Position and skin
Electrode placement
Recording
The trace and finishing
0 of 27 lines tickedRunning total0/32
0 of 27 ticked, 0 of 32 marks

Communication 0/4Systematic 0/11Clinical reasoning 0/2Professionalism 0/3Patient safety 0/12

Your ticks stay on this device, ready for when you come back.

How it is marked

What this station is really testing

An ECG takes two minutes to record and is acted on for days, so the examiner is listening for the things that make a trace true: the right patient's name on it, ten electrodes in the right places and a baseline clean enough to read. Almost every error in this station produces a trace that still looks like an ECG. Chest electrodes put on by eye sit too high, and the result can pass for an old infarct. Limb electrodes moved to the shoulders and hips change the axis. A machine that still holds the last patient's details prints a perfectly good recording under somebody else's name. None of this is visible to the doctor who reads the trace later, which is why the marks are for the method and not for getting paper out of the machine.

Read the rest

This is a talk-through, so say where your fingers go. Find the sternal angle, slide into the second space and count down to the fourth: V1 at the right sternal edge, V2 at the left, V4 in the fifth space in the mid-clavicular line, V3 halfway between them, then V5 and V6 level with V4 in the anterior and mid-axillary lines. Say that the limb electrodes go just above the wrists and ankles. Say twenty-five millimetres a second and ten millimetres per millivolt. Then expect the first recording to be poor. The answer is on the bed and not on the machine: look at Mrs Ashworth, find out why she cannot keep still, put that right and record again. Reaching for the filter first is the wrong reflex, because it tidies the trace by distorting it.

Before the trace leaves the bedside, read the name printed on it. Write on it whether she had symptoms while it was recorded, then take it to someone who can read it. In Year 3, a correctly identified patient, electrodes placed by landmark, a repeated recording after the poor one and a trace carrying her details is the standard. In Year 4, check the contacts before you repeat, explain why the filter stays off, annotate the trace and say how you would spot swapped arm leads. In Year 5, be able to say what misplaced electrodes do to the complexes, when posterior or right-sided leads are recorded, and how quickly a trace must reach a reader when the patient has chest pain.

How practical skills stations are marked, step by step

Where the 32 marks are at Year 3

  • Communication4
  • Systematic approach11
  • Clinical reasoning2
  • Professionalism3
  • Patient safety12

2 critical lines. Miss one and the station is failed, whatever the total.

What a pass looks like in each year

The notice on the door reads the same in Year 3 and in Year 5. What changes is the sheet behind it.

Year 3

27 lines, 32 marks

The sheet starts here.

Records a readable trace from correctly placed electrodes on a correctly identified patient, deals with the poor first recording at the bedside, and checks the name on the paper without being asked.

What lifts it to a clear pass

  • Learn the chest positions as a route for your finger: sternal angle, second space, count to four, right edge, left edge, fifth space mid-clavicular, halfway, then round at the same level.
  • Say V4 before V3. It shows you know why.
  • When the trace is fuzzy, your next sentence is about the patient, not the machine.
  • Make reading the name on the printout a habit that happens before the electrodes come off.

Year 4

32 lines, 37 marks

Adds five lines, most of them on systematic approach and clinical reasoning.

Explains the reasons behind the method: why the filter stays off, why limb electrodes stay on the limbs, what goes on the trace by hand, and how a cable swap shows itself.

What lifts it to a clear pass

  • Know the three kinds of interference by their look: thick and fuzzy for muscle, wandering for movement or a loose electrode, fine and regular for mains.
  • Before repeating, press each electrode and follow each cable. Say that you are doing it.
  • Write "no chest pain at time of recording" on the trace. The reader will look for it.
  • Negative P, QRS and T in lead I with normal chest leads is swapped arm cables until proved otherwise.

Year 5

34 lines, 39 marks

Adds two lines, all on clinical reasoning.

Treats the recording as the first step of a clinical decision: knows how technique falsifies a trace, when to add leads, and how fast a trace must be seen when the patient has symptoms.

What lifts it to a clear pass

  • Be able to say what V1 and V2 two spaces too high look like, and which diagnoses that imitates.
  • Know the posterior positions: V7, V8 and V9 at the level of V4, in the posterior axillary line, the mid-scapular line and at the border of the spine.
  • In suspected acute coronary syndrome the ECG is recorded as soon as possible and put in a reader's hand, not in the notes.
  • Compare with a previous trace whenever one exists, and say so.

A system you can run under pressure

Learn the sequence, not the station. A fixed order is what stops you dropping a step when the clock is running and the examiner is writing.

Patient and machine

  • Ask her to state her full name and date of birth and check them against the wristband — at least two identifiers, because the trace will be filed and compared for years
  • Look at the machine before you look at the patient: cables intact, paper loaded, battery charged, date and time right — a fault found now costs nothing
  • Clear the last patient's details and enter hers — some machines keep the previous name, and "copy" can print the previous patient's trace
  • Explain in plain words: stickers on the chest, arms and legs, painless, a few minutes, lying still — a patient who knows it cannot hurt relaxes, and a relaxed patient gives a clean trace
  • Offer a chaperone, draw the curtains and ask whether she has chest pain, palpitations or dizziness now — symptoms at the time of recording belong on the trace

Position and skin

  • Sit her back at about 45 degrees with arms and legs supported by the bed — tense or unsupported limbs put muscle signal on top of the heart's
  • Expose the chest to the waist, and cover her again once the cables are on — you cannot place by landmark through clothing, and she should not lie uncovered while you work
  • Wipe off lotion or sweat, dry the skin and rub it lightly with gauze — the electrode needs dry skin to stick to and to conduct
  • Remove hair only where an electrode will not stick, with her consent and a single-use razor — and the blade goes in the sharps bin
  • Check the electrodes are in date and the gel is moist — a dried-out electrode gives a wandering baseline

Ten electrodes

  • Limb electrodes on the forearms just above the wrists and on the lower legs just above the ankles — on the trunk they change the axis and the voltages
  • Find the sternal angle, slide into the second intercostal space and count down to the fourth — V1 and V2 put on by eye are usually a space or two too high
  • V1 in the fourth space at the right sternal edge, V2 in the fourth space at the left — count the left side separately, because the spaces do not always line up
  • V4 in the fifth space in the mid-clavicular line, then V3 midway between V2 and V4 — V4 goes on first because V3 has no landmark of its own
  • V5 in the anterior axillary line and V6 in the mid-axillary line, both level with V4 — horizontal, not following the rib as it curves upwards
  • Where her breast lies over the V4 to V6 positions, lift it with the back of your hand, with her permission, and place those electrodes beneath it at the same landmarks and level, not lower — that is the convention, and it keeps serial traces comparable

Recording, and when the trace is poor

  • Check 25 mm/s and 10 mm/mV with the filter off, ask her to lie still and breathe normally, and press start — every measurement on an ECG assumes those two numbers
  • Look at the trace before you take anything off: a steady baseline and P waves, QRS complexes and T waves you can see in every lead — otherwise it is not finished
  • If the baseline is fuzzy, look at the patient first: cold, shivering, tense, in pain or with a tremor — muscle artefact is cured at the bedside
  • Warm and reassure her, let her hands go loose, check each electrode and cable, and record again — an unreadable trace handed on is a delay disguised as a result
  • Use the filter only when the interference cannot be removed, and write that on the trace — it smooths the baseline by distorting the ECG

The paper

  • Read the name, date of birth and number printed on the trace before it leaves the bedside — this is the check that catches the last patient's details
  • Write on it any symptoms during the recording and anything that was not standard — a reader cannot allow for what they have not been told
  • Take it to a doctor who can read it now, and say why it was done — an ECG in a set of notes has been recorded but not seen
  • Remove the electrodes, help her dress, bin the electrodes as clinical waste and clean your hands — then wipe the cables for the next patient

The lines that carry marks

Model lines for the moments this station turns on. Make them your own: examiners can hear a recited script.
  • Confirming identity

    Can you tell me your full name and your date of birth, please? ... Thank you, I am checking that against your wristband.
  • Explaining

    This is a tracing of your heart. I will put ten sticky pads on your chest, arms and legs and clip some wires to them. It does not hurt and it takes a few minutes. I will need you to slip your gown down to your waist. Is that all right?
  • The machine

    The screen still has the last patient's name on it, so I am clearing that and entering Mrs Ashworth's name, date of birth and hospital number before I record anything.
  • Finding the fourth space

    I feel for the ridge of the sternal angle, slide across to the second space on her right and count down: second, third, fourth. V1 goes here at the edge of the sternum, and I count again on the left for V2.
Show the other 5
  • Under the breast

    I need to put three of these pads just under your left breast. May I lift it slightly with the back of my hand?
  • Before pressing start

    Twenty-five millimetres a second, ten millimetres per millivolt, filter off. Try to let your arms go floppy and breathe normally for me. No need to hold your breath.
  • After the poor trace

    That trace is too fuzzy to read, so I am not going to use it. You are shivering. Let me get a blanket over you and give you a minute to warm up, then I will check the pads and record it again.
  • Checking the paper

    Before I take anything off I am reading the top of the trace: Brenda Ashworth, her date of birth, her number, the right date and time. I am writing "no chest pain at time of recording" on it.
  • Handing it on

    This is Mrs Ashworth's ECG, recorded just now for her dizzy spell. She had no symptoms while I recorded it. Could you look at it now, please?

How candidates lose marks here

  • Placing V1 and V2 by eye — count down from the sternal angle, because electrodes placed by eye sit a space or two too high and can make a healthy heart look infarcted.
  • Recording under the last patient's name — look at the machine's screen before you start and at the printout before you leave.
  • Switching the filter on for a fuzzy trace — look at the patient first, because a blanket fixes shivering and the filter only hides it.
  • Putting the limb electrodes on the shoulders and hips because it is quicker — they go above the wrists and ankles, or the axis and voltages change.
  • Handing over the first trace with an apology — a trace you cannot read is not a result, so repeat it.
  • Letting V5 and V6 climb with the rib — keep them on a horizontal line with V4.
  • Leaving her uncovered while you sort out the machine — cover her once the cables are on. It is her dignity, and in this station it is also the trace.

Now sit it live

An AI examiner runs the station by voice. Say each step out loud, from checks and consent to disposal, and they tell you what happens. It rehearses your sequence and your safety checks. It cannot see your hands, so practise the technique itself on a manikin.1 credit, £3.67–£6 a station. No subscription, and credits never expire.
AI OSCE station · UKReal screen
A live AI OSCE skills station from the UK library, peripheral cannulation: the candidate says each step and an examiner tells them what happens. The desk shows the kit on the trolley and the candidate's notes.
AI OSCE station · UKReal screen
The candidate's desk on a phone during an AI OSCE knee examination from the UK library: the instructions, and the observations the examiner has just handed over.
A real screen of the live room on a skills station: you say each step and the examiner tells you what happens.