Practical skills · Procedures
Recording a 12-lead ECG OSCE mark scheme, early clinical
Record a 12-lead ECG on a woman admitted after a dizzy spell, talked through to the examiner: who she is, where each of the ten electrodes goes, what the machine is set to, what you do when the first trace cannot be read, and whose name is on the paper.
Early clinical is where this sheet starts: 27 lines and 32 marks, most of them on systematic approach and patient safety.
- 1:00 + 8:00 reading, then the room
- 5 examiner questions
- Sat from early clinical
- Free to practise, no account needed
- Reviewed


How do you want to practise?
Read the notice, run the clock, then mark yourself on the real sheet.
You present to a friend, who runs the clock and marks on their own phone.
Their marks land on your screen.
An AI examiner runs the station by voice. It cannot see your hands, so practise the technique itself on a manikin.
Candidate instructions
The notice on the door
Mrs Lorraine Petersen, 72, was brought to the emergency department an hour ago after a dizzy spell while waiting for a bus. The resident looking after her has asked for a 12-lead ECG. Please record it, talking the examiner through each thing you check, say and do, and why, including exactly where each electrode goes. There is nothing to touch: the examiner will tell you what you see and what the patient says. You have eight minutes. Towards the end the examiner will ask you some questions about recording an ECG.
1 min reading, then 8 min in the station
On the trolley
- ECG machine with a ten-wire patient cable
- Disposable tab electrodes in a foil pack
- Skin-cleansing wipes and gauze swabs
- Single-use razor
- Sheet to cover the patient
- Alcohol-based hand rub
- Pen
- Sharps container and clinical waste bag
Your notes from reading time
Saved in this browser only.
Early clinical is where this sheet starts: 27 lines and 32 marks, most of them on systematic approach and patient safety.
Early clinical standard. Records a readable trace from correctly placed electrodes on a correctly identified patient, deals with the poor first recording at the bedside, and checks the name on the paper without being asked.
2 critical lines. Miss one and the station is failed, however well the rest goes.
Communication 0/4Systematic 0/11Clinical reasoning 0/2Professionalism 0/3Patient safety 0/12
Run the station before you look
The answers are the most useful thing on this page once you have had a go, and the least useful before.
When the candidate asks the patient to state her name and date of birth
She says Lorraine Petersen, the fourth of February nineteen fifty-four.When the candidate only asks "Are you Lorraine Petersen?", or reads her name out to her for a yes
She says yes.When the candidate checks the ID band
The ID band reads Lorraine Petersen, four two nineteen fifty-four, medical record number three one eight six five two.ID band: PETERSEN, Lorraine | DOB 04/02/1954 | MRN 318652
When the candidate explains the ECG and asks for her consent
She says yes, that is fine.When the candidate offers a chaperone
She says she does not need anyone, thank you.When the candidate asks whether she has chest pain, palpitations or dizziness now
She says she has no chest pain, no palpitations and does not feel dizzy now.When the candidate asks whether she is comfortable or warm enough, before the first recording
She says she is a bit chilly.When the candidate checks the ECG machine or looks at its screen
The cables are intact and there is paper. The screen shows the name Arthur Pell.When the candidate enters her details into the machine
Her name, date of birth and medical record number are now on the screen.When the candidate asks her to undress to the waist or exposes her chest
She slips her gown down to her waist.When the candidate inspects the skin of her chest
The skin is intact, with no hair. It feels greasy with body lotion.When the candidate checks the electrodes are in date or that the gel is moist
The pack is in date and the gel is moist.When the candidate asks her permission to lift her breast or to place electrodes under it
She says yes, go ahead.When the candidate checks the paper speed, the gain or the filter on the machine
The screen reads twenty-five millimetres per second, ten millimetres per millivolt, filter off.When the candidate presses start to record the first trace
The trace prints. The baseline is thick and jagged in every lead, and you cannot make out the P waves.When the candidate looks at the patient or asks how she is, after the poor trace
She is shivering and her fists are clenched. She says she has been cold since she came in.When the candidate checks the electrodes or the cable connections, after the poor trace
All ten electrodes are stuck down and every cable is attached.When the candidate warms her, adds a blanket or waits for her to settle, after the poor trace
After a minute the shivering stops and her hands relax.When the candidate switches the filter on and records while she is still shivering
The trace prints. The baseline is smoother but still uneven.When the candidate records again, without switching the filter on and without doing anything about her shivering
The trace prints with the same thick, jagged baseline.When the candidate records again once she has been warmed or has settled
The trace prints with a steady baseline and clear waves in all twelve leads.When the candidate checks the patient details printed on the trace, having entered hers into the machine
The trace is headed Lorraine Petersen, with her date of birth, her medical record number and the time now.When the candidate checks the patient details printed on the trace, without having entered hers into the machine
The trace is headed Arthur Pell.When the candidate takes or shows the trace to the doctor
The doctor takes it from you.When the candidate asks her how she is at the end
She says she is fine, and warmer now.
How would you recognise from the trace that the two arm leads had been swapped? · Mid clinical
Lead I gives it away. Lead I looks from the right arm towards the left, so swapping the two arm cables turns it upside down: the P wave, the QRS complex and the T wave are all negative in lead I, which almost never happens in a normal heart. Leads aVR and aVL change places, so aVR becomes upright, and leads II and III change places as well. The chest leads are not affected, so the R wave still grows normally from V1 to V6. That is how I tell it from dextrocardia, which also inverts lead I but has R waves that shrink across the chest. If I saw this I would check the cables by the letters on each one and by colour, which on machines that follow the international code is red on the right arm, yellow on the left arm, green on the left leg and black on the right leg, and record again. Swaps that involve a leg cable are harder to see. One clue is a limb lead that is almost a flat line, which happens when an arm cable and the right leg cable are exchanged. Comparing with an earlier trace helps, and I would never hand on a trace I thought was wrongly connected.
What happens to the trace if V1 and V2 are placed too high, or if the limb electrodes are placed on the trunk? · Final year
Both errors give a trace that is abnormal in a believable way. V1 and V2 in the second or third space, which is where they end up when they are placed by eye, look down on the heart from above. The R wave in V1 and V2 gets smaller, an rSr pattern can appear and the P wave in V2 can turn negative, so the trace can be read as an old anteroseptal infarct with poor R wave progression, as incomplete right bundle branch block, or even as a Brugada pattern. Limb electrodes on the shoulders and hips shift the axis to the right and change the height of the limb lead complexes, and inferior Q waves can shrink or disappear, so an old inferior infarct can be hidden. There is a second cost, which is comparison: serial ECGs can only be compared if the electrodes were in the same places each time, and a millimetre of ST change matters. So I place by landmark from the sternal angle every time. If I cannot use a standard position, because of a dressing, an amputation or a patient who has to stay in a wheelchair, I write on the trace exactly what I did.
Suppose that, as you attach the electrodes, she tells you she has a heavy pain in the centre of her chest. What changes? · Final year
The ECG becomes urgent, and I am the person who has to raise the alarm. I would tell her I am getting help, call the nurse and the resident straight away, and carry on recording, because a trace taken during pain is the most useful one she will have. The Australian acute coronary syndromes guideline asks for an ECG within 10 minutes of first clinical contact, read by a clinician experienced in interpreting ECGs. I would not cut corners on placement, because a clean trace from the wrong positions misleads. I write "chest pain at time of recording" and the time on it, and I put it into the hand of a doctor who can read it now. While that happens she needs observations, monitoring and an assessment from airway to exposure, and I would stay with her. If her observations meet the calling criteria on the unit's chart, that is a MET call. I would expect to be asked to repeat the ECG if the pain continues or changes, and possibly for extra leads. With ST depression in V1 to V3 that means posterior leads: V4, V5 and V6 are moved round to the posterior axillary line, the mid-scapular line and the border of the spine, at the same level, and relabelled V7, V8 and V9. With an inferior infarct it means a right-sided lead, V4R, in the mirror image of V4. Every moved lead is relabelled by hand on the trace.
Suppose a trace prints with a thick, fuzzy baseline in every lead. What can cause that, and what do you do before you record again? · Early clinical
That is muscle artefact: the electrodes are picking up skeletal muscle as well as the heart. The usual reasons are a patient who is cold and shivering, tense, in pain or holding their arms stiff, and a tremor such as Parkinson's disease. So I look at the patient before I touch the machine. I would cover and warm her, make sure her arms and legs are supported by the bed, ask her to let her hands go loose, and wait until she has settled. The other kinds of interference look different. A baseline that wanders up and down comes from movement, deep breathing or an electrode that is not stuck to clean, dry skin. A fine, perfectly regular fuzz is mains interference from electrical equipment nearby. So I also press each electrode down and follow each cable to its connection. Then I record again with the filter still off. Only if the tremor cannot be removed do I switch the filter on, and I write that on the trace, because the filter distorts the recording as well as smoothing it. I do not hand over a trace I cannot read.
What are the standard paper speed and gain for a 12-lead ECG, and what does that make one small square worth? · Early clinical
Twenty-five millimetres per second and ten millimetres per millivolt. At that speed one small square, which is one millimetre, is 0.04 seconds and one large square is 0.2 seconds, so five large squares pass every second and three hundred every minute. That is where the rate rule comes from: three hundred divided by the number of large squares between two R waves. At that gain one millivolt is ten millimetres, two large squares tall, so each small square is 0.1 millivolts. Most machines print a calibration mark on each line, a box one millivolt high, and it should stand ten millimetres tall. I check the mark and the printed settings before I measure anything, because every interval and every voltage criterion assumes them. At fifty millimetres per second everything looks twice as wide and the rate looks half as fast. The gain is sometimes halved to five millimetres per millivolt on purpose, when the complexes are so tall that they overlap, and then ST elevation looks half as big. Any change from the standard settings has to be marked clearly on the trace.
How it is marked
What this station is really testing
An ECG takes two minutes to record and is acted on for days, so the examiner is listening for the things that make a trace true: the right patient's name on it, ten electrodes in the right places and a baseline clean enough to read. Almost every error in this station produces a trace that still looks like an ECG. Chest electrodes put on by eye sit too high, and the result can pass for an old infarct. Limb electrodes moved to the shoulders and hips change the axis. A machine that still holds the last patient's details prints a perfectly good recording under somebody else's name. None of this is visible to the doctor who reads the trace later, which is why the marks are for the method and not for getting paper out of the machine.
Read the restShow less
This is a talk-through, so say where your fingers go. Find the sternal angle, slide into the second space and count down to the fourth: V1 at the right sternal edge, V2 at the left, V4 in the fifth space in the mid-clavicular line, V3 halfway between them, then V5 and V6 level with V4 in the anterior and mid-axillary lines. Say that the limb electrodes go just above the wrists and ankles. Say twenty-five millimetres a second and ten millimetres per millivolt. Then expect the first recording to be poor. The answer is on the bed and not on the machine: look at Mrs Petersen, find out why she cannot keep still, put that right and record again. Reaching for the filter first is the wrong reflex, because it tidies the trace by distorting it.
Before the trace leaves the bedside, read the name printed on it. Write on it whether she had symptoms while it was recorded, then take it to someone who can read it. In the early clinical year, a correctly identified patient, electrodes placed by landmark, a repeated recording after the poor one and a trace carrying her details is the standard. In the mid clinical year, check the contacts before you repeat, explain why the filter stays off, annotate the trace and say how you would spot swapped arm leads. In the final year, be able to say what misplaced electrodes do to the complexes, when posterior or right-sided leads are recorded, and how quickly a trace must reach a reader when the patient has chest pain.
Where the 32 marks are at Early clinical
- Communication4
- Systematic approach11
- Clinical reasoning2
- Professionalism3
- Patient safety12
2 critical lines. Miss one and the station is failed, whatever the total.
What a pass looks like at each stage
Early clinical
27 lines, 32 marks
The sheet starts here.
Records a readable trace from correctly placed electrodes on a correctly identified patient, deals with the poor first recording at the bedside, and checks the name on the paper without being asked.
What lifts it to a clear pass
- Learn the chest positions as a route for your finger: sternal angle, second space, count to four, right edge, left edge, fifth space mid-clavicular, halfway, then round at the same level.
- Say V4 before V3. It shows you know why.
- When the trace is fuzzy, your next sentence is about the patient, not the machine.
- Make reading the name on the printout a habit that happens before the electrodes come off.
Mid clinical
32 lines, 37 marks
Adds five lines, most of them on systematic approach and clinical reasoning.
Explains the reasons behind the method: why the filter stays off, why limb electrodes stay on the limbs, what goes on the trace by hand, and how a cable swap shows itself.
What lifts it to a clear pass
- Know the three kinds of interference by their look: thick and fuzzy for muscle, wandering for movement or a loose electrode, fine and regular for mains.
- Before repeating, press each electrode and follow each cable. Say that you are doing it.
- Write "no chest pain at time of recording" on the trace. The reader will look for it.
- Negative P, QRS and T in lead I with normal chest leads is swapped arm cables until proved otherwise.
Final year
34 lines, 39 marks
Adds two lines, all on clinical reasoning.
Treats the recording as the first step of a clinical decision: knows how technique falsifies a trace, when to add leads, and how fast a trace must be seen when the patient has symptoms.
What lifts it to a clear pass
- Be able to say what V1 and V2 two spaces too high look like, and which diagnoses that imitates.
- Know the posterior positions: V7, V8 and V9 at the level of V4, in the posterior axillary line, the mid-scapular line and at the border of the spine.
- In suspected acute coronary syndrome the ECG is recorded and read within 10 minutes of first clinical contact, so it goes into a reader's hand, not into the notes.
- Compare with a previous trace whenever one exists, and say so.
A system you can run under pressure
Patient and machine
- Ask her to state her full name and date of birth and check them against the ID band — at least two identifiers, because the trace will be filed and compared for years
- Look at the machine before you look at the patient: cables intact, paper loaded, battery charged, date and time right — a fault found now costs nothing
- Clear the last patient's details and enter hers — some machines keep the previous name, and "copy" can print the previous patient's trace
- Explain in plain words: stickers on the chest, arms and legs, painless, a few minutes, lying still — a patient who knows it cannot hurt relaxes, and a relaxed patient gives a clean trace
- Offer a chaperone, draw the curtains and ask whether she has chest pain, palpitations or dizziness now — symptoms at the time of recording belong on the trace
Position and skin
- Sit her back at about 45 degrees with arms and legs supported by the bed — tense or unsupported limbs put muscle signal on top of the heart's
- Expose the chest to the waist, and cover her again once the cables are on — you cannot place by landmark through clothing, and she should not lie uncovered while you work
- Wipe off lotion or sweat, dry the skin and rub it lightly with gauze — the electrode needs dry skin to stick to and to conduct
- Remove hair only where an electrode will not stick, with her consent and a single-use razor — and the blade goes in the sharps container
- Check the electrodes are in date and the gel is moist — a dried-out electrode gives a wandering baseline
Ten electrodes
- Limb electrodes on the forearms just above the wrists and on the lower legs just above the ankles — on the trunk they change the axis and the voltages
- Find the sternal angle, slide into the second intercostal space and count down to the fourth — V1 and V2 put on by eye are usually a space or two too high
- V1 in the fourth space at the right sternal edge, V2 in the fourth space at the left — count the left side separately, because the spaces do not always line up
- V4 in the fifth space in the mid-clavicular line, then V3 midway between V2 and V4 — V4 goes on first because V3 has no landmark of its own
- V5 in the anterior axillary line and V6 in the mid-axillary line, both level with V4 — horizontal, not following the rib as it curves upwards
- Where her breast lies over the V4 to V6 positions, lift it with the back of your hand, with her permission, and place those electrodes beneath it at the same landmarks and level, not lower — that is the convention, and it keeps serial traces comparable
Recording, and when the trace is poor
- Check 25 mm/s and 10 mm/mV with the filter off, ask her to lie still and breathe normally, and press start — every measurement on an ECG assumes those two numbers
- Look at the trace before you take anything off: a steady baseline and P waves, QRS complexes and T waves you can see in every lead — otherwise it is not finished
- If the baseline is fuzzy, look at the patient first: cold, shivering, tense, in pain or with a tremor — muscle artefact is cured at the bedside
- Warm and reassure her, let her hands go loose, check each electrode and cable, and record again — an unreadable trace handed on is a delay disguised as a result
- Use the filter only when the interference cannot be removed, and write that on the trace — it smooths the baseline by distorting the ECG
The paper
- Read the name, date of birth and number printed on the trace before it leaves the bedside — this is the check that catches the last patient's details
- Write on it any symptoms during the recording and anything that was not standard — a reader cannot allow for what they have not been told
- Take it to a doctor who can read it now, and say why it was done — an ECG in a set of notes has been recorded but not seen
- Remove the electrodes, help her dress, bin the electrodes as clinical waste and clean your hands — then wipe the cables for the next patient
The lines that carry marks
Confirming identity
Can you tell me your full name and your date of birth, please? ... Thank you, I am checking that against your ID band.
Explaining
This is a tracing of your heart. I will put ten sticky pads on your chest, arms and legs and clip some wires to them. It does not hurt and it takes a few minutes. I will need you to slip your gown down to your waist. Is that all right?
The machine
The screen still has the last patient's name on it, so I am clearing that and entering Mrs Petersen's name, date of birth and medical record number before I record anything.
Finding the fourth space
I feel for the ridge of the sternal angle, slide across to the second space on her right and count down: second, third, fourth. V1 goes here at the edge of the sternum, and I count again on the left for V2.
Show the other 5Show fewer
Under the breast
I need to put three of these pads just under your left breast. May I lift it slightly with the back of my hand?
Before pressing start
Twenty-five millimetres a second, ten millimetres per millivolt, filter off. Try to let your arms go floppy and breathe normally for me. No need to hold your breath.
After the poor trace
That trace is too fuzzy to read, so I am not going to use it. You are shivering. Let me get a blanket over you and give you a minute to warm up, then I will check the pads and record it again.
Checking the paper
Before I take anything off I am reading the top of the trace: Lorraine Petersen, her date of birth, her number, the right date and time. I am writing "no chest pain at time of recording" on it.
Handing it on
This is Mrs Petersen's ECG, recorded just now for her dizzy spell. She had no symptoms while I recorded it. Could you look at it now, please?
How candidates lose marks here
- Placing V1 and V2 by eye — count down from the sternal angle, because electrodes placed by eye sit a space or two too high and can make a healthy heart look infarcted.
- Recording under the last patient's name — look at the machine's screen before you start and at the printout before you leave.
- Switching the filter on for a fuzzy trace — look at the patient first, because a blanket fixes shivering and the filter only hides it.
- Putting the limb electrodes on the shoulders and hips because it is quicker — they go above the wrists and ankles, or the axis and voltages change.
- Handing over the first trace with an apology — a trace you cannot read is not a result, so repeat it.
- Letting V5 and V6 climb with the rib — keep them on a horizontal line with V4.
- Leaving her uncovered while you sort out the machine — cover her once the cables are on. It is her dignity, and in this station it is also the trace.
Where to read more
- Australian clinical guideline for diagnosing and managing acute coronary syndromes 2025National Heart Foundation of Australia and CSANZ · guide
- Acute Coronary Syndromes Clinical Care StandardAustralian Commission on Safety and Quality in Health Care · reference
- Recording a standard 12-lead electrocardiogram: clinical guidelines by consensus (2024)Society for Cardiological Science and Technology · guide
- Recommendations for the standardization and interpretation of the electrocardiogram, part I: the electrocardiogram and its technologyAmerican Heart Association · reference
- Communicating for Safety Standard: correct identification and procedure matchingAustralian Commission on Safety and Quality in Health Care · reference
- Electrocardiogram (ECG)healthdirect · guide